KontrollVahend ja Tehnika (KVT) OÜ
Exclusive representative
Hottinger Baldwin Messtechnik GmbH, Germany
+372 6006190
+372 6006680


Products

Force Transducer C18

Special features

  • Compressive force transducer
  • Nominal (rated) forces10 kN to 5 MN
  • Compact dimensions
  • Low weight
  • Delivery includes force application parts
  • Classification option with DKD calibration certificate to ISO 376: Class 0.5

Type

C18

Data per VDI 2638

Nominal (rated) force

Fnom

kN

10 - 200

300

500 - 1000

2000 - 3000

5000

Class to ISO 376 (0.2 Fnom to Fnom) 1)

0.5

Nominal (rated) sensitivity

Cnom

mV/V

2

Rel. sensitivity error (compression)

dC

%

0.1

Relative zero signal error

ds,o

%

1

Relative zero error (zero signal return) 1)

fo

%

0.012

0.024

Hysteresis error (0.2Fnom to Fnom ) 1)

u

%

0.08

Relative reproducibility and repeatability errors (0.2Fnom to Fnom) for:

a constant mounting position 1)

varying mounting positions 1)

bl b

%

%

0.04

0.08

Non‐linearity

dlin

%

0.05

Effect of temperature on sensitivity, per 10 K

related to nominal (rated) sensitivity

TKc

%

0.01

Effect of temperature on zero signal, per 10 K

related to nominal (rated) sensitivity

TK0

%

0.01

Effect of lateral forces (lateral force 10% Fnom)2)

dQ

%

0.035

0.1

0.15

Effect of eccentricity per mm

dE

%

0.02

Relative creep over 30 min

dcrF+E

%

0.03

Input resistance

Ri

Ω

4450±100

Output resistance

Ro

Ω

4010±2

Insulation resistance

Ris

Ω

>50 x 109

Reference excitation voltage

Uref

V

5

Operating range of excitation voltage

BU,G T

V

5 to 30

Carrier frequency of excitation voltage

Hz

::;600

Nominal (rated) temperature range

Bt,nom

oC

+10 to +40

Operating temperature range

Bt,G

oC

‐30 to +80

Storage temperature range

Bt,S

oC

‐50 to +85

Reference temperature

tref

oC

+22

Max. operating force

(FG)

%

170

150

135

Limit force

(FL)

%

170

150

135

Breaking force

(FB)

%

400

320

290

Static lateral limit force2)

(FQ)

%

0.27⋅Fnom; (to Fz ≤ 0.5⋅Fnom) 0.45⋅(Fnom‐0.8⋅Fz); (for Fz > 0.5⋅Fnom) (Fz = force in direction of measurement)

Force Transducer C18 - datasheet PDF, 0.09 Kb