Special features

  • Symmetrical structure for increased stability
  • Nominal (rated) forces: 20 kN,50 kN, 100 kN, 140 kN, 190 kN,330 kN and 700 kN
  • High cut‐off frequency
  • Robust charge cable connection

Piezoelectric force washer

CFW / ...

Nominal (rated) force

kN

20

50

100

140

190

330

700

Sensitivity (typical)1)

pC/N

-4

-4.3

Relative reversibility error

% of E.

<1

Relative linearity error 2)

% of E.

<1

Crosstalk3)

N/N

0.03

0.03

0.04

0.03

0.03

0.03

0.03

from Fx,y to Fz

from Mx,y to Fz

N/Nm

0.002

0.002

0.002

0.002

0.003

0.003

0.003

Max. operating force

% of Fnom

120

Maximum bending moment with 3)

Nm

0

0

0

0

0

0

0

with Fz = 0%

with Fz = 50%

21

80

220

395

650

1200

4000

with Fz = 100%

0

0

0

0

0

0

0

Max. permissible lateral force 4)

% Fz

10

Breaking force

% of Fnom

200

Insulation resistance

W

> 1013

Nominal (rated) temperature range

°C

-40 ¼ +120

Operating temperature range

°C

-40 ¼ +120

Storage temperature range

°C

-40 ¼ +120

Nominal (rated) displacement

mm

2

3.5

5.8

7.5

9.4

13.5

28

Fundamental resonance frequency

kHz

60

55

45

35

35

30

30

Permissible vibrational stress

(vibration bandwidth)

% Fnom

100 for compressive force

Weight

g

7

22

37

57

78

155

365

Degree of protection per DIN EN 60529

(with connected charge cable)

IP65

Connection

10‐32 UNF

Specifications (data per VDI/VDE 2638 standards)

1) Must be calibrated under mounting conditions

2) Pre‐stressed, typically 0.5% of full scale

3) Fz is the force in the measurement direction

4) related to a point of contact on the force application surface

Force Transducer PaceLine CFW - datasheet PDF, 0.16 Kb