KontrollVahend ja Tehnika (KVT) OÜ
Exclusive representative
Hottinger Baldwin Messtechnik GmbH, Germany
+372 6006190
+372 6006680


Products

Force Transducer U15

Special features

  • Tensile/compressive force transducer
  • Nominal (rated) forces 2.5 kN to 1 MN
  • Class 0.5 to ISO 376 (in combination with DKD calibration certificate)
  • Electronic bending moment adjustment
  • Double bridge version option

Type

U15

Data per VDI 2638 and ISO 376

Nominal (rated) force

Fnom

kN

2.5

5

10

25

50

100

250

500

MN

1

Class to ISO 376 (0.2 Fnom to Fnom) 1)

0.5

Nominal (rated) sensitivity

Rel. zero signal error

Cnom

ds,o

mV/V

%

2 … 33)

4 … 4,84)

<±1

Rel. reproducibility and repeatability errors (0.2Fnom to Fnom) for:

a constant mounting position

varying mounting positions

Rel. interpolation error (0.2Fnom to Fnom) Relative zero error (zero signal return) Rel. reversibility error (0.2Fnom to Fnom)

b' b fc fo ν

%

%

%

%

%

<±0.025

<±0.05

<±0.01 <±0.04

<±0.01

<±0.05

<±0.02

<±0.15

<±0.075

<±0.1

<±0.125

Relative linearity error

dlin

%

<±0.03

<±0.04

<±0.06

Temperature effect on sensitivity per 10 K, relative to the nominal (rated) sensitivity

Temperature effect on the zero signal per 10 K,

related to the nominal (rated) sensitivity

TKc

TK0

%

%

<±0.015

<±0.01

Relative creep over 30 min

dcrF+E

%

<±0.04

<±0.025

Effect of lateral forces (lateral force 10% Fnom)

dQ

%

<0.015

Input resistance Output resistance Insulation resistance

Ri

Ro

Ris

Ω

Ω

Ω

>345

220 to 300

>2 x 109

Reference excitation voltage Operating range of excitation voltage

Uref

BU,G

V

V

5

0.5 to 12

Nominal temperature range Operating temperature range Storage temperature range Reference temperature

Bt,nom

Bt,G Bt,S

tref

°C

°C

°C

°C

+10 to +40

-30 to +85

-30 to +85

+22

Max. operating force Breaking force

(FG)

(FB)

%

%

115

200

Limit torque

(MG)

N⋅m

15

30

60

155

180

635

1320

2855

5715

Nominal (rated) displacement Fundamental resonance frequency Rel. permissible vibrational stress

Snom fG

Frb

mm

kHz

%

0.04

2.7 3.8 5.6

0.06

5.3 7.5 4.3

0.08

5.8

0.1

4.9

0.12

4.0

100

Weight

kg

1.4

3.3

10.5

27

73

Degree of protection per EN 60529

IP67 2)

Plug connection, six‐wire configuration

Bayonet or thread (option) plug‐compatible to MIL‐C‐26482 series 1

Transducer identification (option)

TEDS, per IEEE1451.4

Force Transducer U15 - datasheet PDF, 0.31 Kb