Force Transducer U1A

Special features

  • Tensile force / compressive force transducer
  • Nominal forces 10 N ... 50 N
  • Integrated overload protection
  • Side-force support
  • Variable force introduction

Type

U1A

Accuracy class

0.1

Nominal force

N

10

20

50

Nominal sensitivity

rel. sensitivity deviation

rel. tensile-compressive sensitivity difference rel. deviation from zero

Rel. range of inversion (0.2Fnom to Fnom) Linearity deviation

Cnom dc

dzd ds,o u

dlin

mV/V

%

%

%

%

%

2

0.2

0.4

1

0.15

0.1

Temperature effect on the sensitivity/10 K

by reference to nominal sensitivity

Temperature effect on the zero signal/10 K

by reference to nominal sensitivity

Effect of eccentricity / mm Rel. creep over 30 min

TKC

TK0

dE dcrF+E

%

%

%

%

0.1

0.05

0.1

±0.06

Input resistance

Output resistance Isolation resistance

Re

Ra Ris

>345

300 − 400

>5 x 109

Reference excitation voltage

Operating range of the excitation voltage

Uref

BU,G

V

V

5

0.5 ... 6

Nominal temperature range

Operating temperature range Storage temperature range Reference temperature

Bt,nom

Bt,G Bt,S

tref

oC

oC oC oC

−10 ... +50

−20 ... +50

−30 ... +60

+23

Max. operational force

Limit force Breaking force

Static lateral limit force

(FG)

(FL)

(FB)

(FQ)

%

%

%

%

120

1000

2000

200

500

1000

100

200

400

40

Nominal displacement

Fundamental resonance frequency Weight

Rel. permissible vibrational stress Degree of protection to DIN EN 60529

Snom fG

Frb

mm

Hz kg

%

0.28

300

0.21

450

0.2

700

0.3

70

IP 40

Force Transducer U1A - datasheet PDF, 0.17 Kb