General specifications | ||
Inputs |
Number |
16, electrically isolated from each other and from the supply voltage |
Transducer technologies |
|
Voltage, current, current−fed piezoelectric sensors (IEPE) |
A/D converter |
|
24 Bit Delta Sigma converter |
Data rates |
Hz |
0.1 … 19,200 adjustable for each channel |
Active low-pass filter (Bessel/Butterworth, can be switched off) |
Hz |
0.01 ... 3,000 (−3 dB) |
Transducer identification (TEDS, IEEE 1451.4) max. distance of the TEDS module |
m |
100 |
Transducer connection |
|
Phönix Contact FMC−1,5/8−ST−3,5−RF plug terminal |
Supply voltage range (DC) |
V |
10 ... 30 (24 V nominal (rated) voltage) |
Supply voltage interruption |
|
max. 5 ms at 24 V |
Power consumption |
W |
< 10 |
without adjustable transducer supply | ||
with adjustable transducer supply |
W |
< 13 |
Transducer Excitation (active transducers) |
V |
5 ... 24; adjustable for each channel |
Only channel 1... 8: | ||
Adjustable supply voltage (DC) | ||
Maximum output power |
W |
0.7 each channel / a total of 2 |
Only channel 9 ... 16: Supply voltage (DC) Maximum output current |
V mA |
9 ... 29; Supply voltage of the module −1 V 30 each channel / a total of 75 |
Ethernet (data link) Protocol/addressing Connection Max. cable length to module |
− − m |
10Base-T / 100Base-TX TCP/IP (direct IP address or DHCP) 8P8C plug (RJ-45) with twisted pair cable (CAT-5) 100 |
IEEE1394b FireWire (module synchronization, data link, |
MBaud |
IEEE 1394b (HBM modules only) |
optional supply voltage) |
| |
Baud rate |
400 (approx. 50 MByte/s) | |
Max. current from module to module |
A |
1.5 |
Max. cable length between the nodes |
m |
5 |
Max. number of modules connected in series (daisy chain) |
− |
12 (=11 Hops) |
Max. number of modules in a IEEE1394b FireWire system |
|
|
(including hubs, backplane) |
− |
24 |
Max. number of hops |
− |
14 |
Synchronization options EtherCAT IRIG−B (B000 to B007; B120 to B127) IEEE1588v2 (PTP), NTP |
|
FireWire (automatically, recommended) via CX27 via MX440A- or MX840A input channel via Ethernet |
Nominal (rated) temperature range |
°C [°F] |
−20 ... +65 [−4 ... +149] |
Storage temperature range |
°C [°F] |
−40 ... +75 [−40 ... +167] |
Rel. humidity |
% |
5 ... 95 (non condensing) |
Protection class |
|
III |
Degree of protection |
|
IP20 per EN 60529 |
Mechanical tests |
m/s2 |
50 |
Vibration (30 min) | ||
Shock (6 ms) |
m/s2 |
350 |
EMC requirements |
|
per EN 61326−1 |
Max. input voltage at transducer socket to ground |
V |
transient free |
Pin 4 (TEDS) |
+ 5 | |
Pin 1 (voltage) |
V |
±60 |
Pin 3 (current) |
V |
±1,5 |
Pin 5 (control lead) |
V |
+ 3,3 |
Dimensions, horizontal (W x H x D) |
mm |
52.5 x 200 x 122 (with case protection) 44 x 174 x 119 (without case protection) |
Weight, approx. |
g |
980 |