KontrollVahend ja Tehnika (KVT) OÜ
Exclusive representative
Hottinger Baldwin Messtechnik GmbH, Germany
+372 6006190
+372 6006680


Products

Torque Transducer T22

Special features

  • Nominal (rated) torques 5 N·m, 10 N·m, 20 N·m, 50 N·m, 100 N·m, 200 N·m, 500 N·m and 1 kN·m
  • Nominal (rated) rotational speeds of 9000 rpm to 16 000 rpm
  • Accuracy class: 0.5
  • Contactless transmission of measurement signals
  • Measurement on rotating or stationary parts
  • Cylindrical shaft ends for non-play friction joints
  • ±5 V and 10±8 mA torque output signal

Type

T22

Accuracy class

0.5

Torque measuring system

Nominal (rated) torque Mnom

N×m

5

10

20

50

100

200

500

kN×m

1

Nominal (rated) sensitivity (span between

V

5

torque = zero and nominal (rated) torque Mnom)

Voltage output

Current output

mA

8

Sensitivity tolerance (deviation of the actual output

quantity at Mnom from the nominal (rated) sensitivity)

Voltage output

%

±0.5

Current output

%

±0.5

Output signal at torque = zero

V

0±0.2

Voltage output

Current output

mA

10±0.2

Nominal (rated) output signal

Voltage output

at positive nominal (rated) torque at negative nominal (rated) torque

Current output

at positive nominal (rated) torque at negative nominal (rated) torque

Load resistance (Voltage output)

Burden (current output)

with UB = 12 V with UB = 24 V

Longterm drift over 48h

Voltage output Current output

Cut-off frequency (-3 dB) (volt. output / current output) Group delay (voltage output / current output) Residual ripple

Voltage output

Current output

V V

mA mA

MW

W

W

mV

f!A kHz f!s

mVpp mApp

+5

-5

+18

+2

>1

250

500

<±50

<±80

1

450

<100

<0.1

Effect of temperature per 10 K in the

%

::;±0.2

nominal (rated) temperature range

on the output signal, related to the actual value of the

signal span

on the zero signal, relat. to the nominal (rated) sensitivity

%

::;±0.5

Power supply

V (DC)

11.5 ... 30

Nominal (rated) supply voltage range

Current consumption in measuring mode

A

< 0.2

Nominal (rated) power consumption

W

< 2.4

Permissible residual ripple of supply voltage

mVPP

200

Linearity error including hysteresis, related to the

%

::;±0.3

nominal (rated) sensitivity

Rel. standard deviation of repeatability per DIN

1319, related to the variation of the output signal

%

::;±0.1

Max. level control range1)

voltage output / current output

%

::;120