For the 100% calibration version (standard version) | |||||||||||
Type |
C10 | ||||||||||
Nominal (rated) force |
Fnom |
kN |
2.5 |
5 |
10 |
25 |
50 |
100 |
250 |
500 |
1000 |
Accuracy 1) | |||||||||||
Accuracy class |
0.03 |
0.04 |
0.05 |
0.06 | |||||||
Relative reproducibility and repeatability errors without rotation |
br,g |
% |
0.025 | ||||||||
Hysteresis error (0.4 Fnom) |
u0.4 |
%vI |
0.075 |
0.1 |
0.125 |
0.15 | |||||
%vc |
0.03 |
0.04 |
0.05 |
0.06 | |||||||
Non-linearity |
dlin |
% |
0.03 |
0.04 |
0.06 | ||||||
Relative creep over 30 min |
dcr, F+E |
% |
0.04 |
0.025 | |||||||
Effect of eccentricity |
dE |
%/mm |
0.04 | ||||||||
Temperature influence on sensitivity |
TCC |
%/10K |
0.015 | ||||||||
Effect of temperature on the zero signal |
TC0 |
%/10K |
0.0075 | ||||||||
Electrical values | |||||||||||
Nominal (rated) sensitivity |
Cnom |
mV/V |
2 |
4 | |||||||
Relative zero signal error |
ds,0 |
% |
1 | ||||||||
Relative sensitivity error 1) |
dc |
% |
0.1 | ||||||||
Input resistance |
Ri |
Ω |
>345 | ||||||||
Output resistance |
Ro |
Ω |
280…360 | ||||||||
Insulation resistance |
Ri |
Giga Ω |
>2 | ||||||||
Operating range of excitation voltage |
BU, G |
V |
0.5…12 | ||||||||
Reference excitation voltage |
Uref |
V |
5 | ||||||||
Connector |
Six wire circuit | ||||||||||
Temperature | |||||||||||
Reference temperature |
Tref |
°C |
23 | ||||||||
Nominal (rated) temperature range |
BT, nom |
°C |
-10…+45 | ||||||||
Operating temperature range |
BT, G |
°C |
-30…+85 | ||||||||
Storage temperature range |
BT, S |
°C |
-30…+85 | ||||||||
Mechanical quantities | |||||||||||
Max. operating force |
FG |
% |
120 | ||||||||
Limit force |
FL |
% |
120 | ||||||||
Breaking force |
FB |
% |
>200 | ||||||||
Max. eccentricity |
eG |
mm |
10.2 |
9.9 |
9.1 |
14.1 |
12 |
20.6 |
23.9 | ||
Nominal (rated) displacement |
snom |
mm |
0.04 |
0.06 |
0.08 |
0.1 |
0.12 | ||||
Fundamental resonance frequency |
fG |
kHz |
4.7 |
6.5 |
8.6 |
5.8 |
8.2 |
5.7 |
7.3 |
5.9 |
5.4 |
Relative permissible oscillatory stress |
Frb |
% |
100 | ||||||||
General information | |||||||||||
Degree of protection per DIN 60529 with bayonet connector |
IP67 | ||||||||||
with threaded connector |
IP64 | ||||||||||
with a fixed cable |
IP67 |
IP68 | |||||||||
Measuring body material |
Aluminum |
Stainless steel | |||||||||
Cable 2) |
Measurement cable with TPE insulation, wires twisted in pairs, 6 or 15 m | ||||||||||
Mass |
Force Transducer C10 - datasheet | PDF, 0.17 Kb |