Type |
CFT/... | |||||
Nominal (rated) force |
kN |
5 |
20 |
50 |
70 |
120 |
Sensitivity |
pC/N |
−7,7 |
−7,7 |
−4,1 |
−4,1 |
−4,0 |
Perm. lateral force1) |
% Fnom |
0.5 |
3.5 |
4.8 | ||
Permissible lateral force bei Fz = 0% Max. operational force |
N
kN |
80
5.5 |
160
22 |
1000
60 |
1800
84 |
5800
144 |
Permissible bending moment at Fz = 0% at Fz = 100 % |
Nm Nm |
2 0.5 |
4 2 |
75 20 |
150 20 |
650 250 |
Crosstallk from Fx, y to Fz from Mx, y to Fz |
N/N N/Nm |
< 0.06 −8 |
< 0.05 −6 |
< 0.032 < 0,3 |
< 0.045 < 0,3 |
< 0.08 < 0,25 |
Breaking force |
kN |
10 |
31 |
160 |
220 |
510 |
Natural frequency |
kHz |
40 |
36 |
54 |
46 |
31 |
Oscillation width |
% Fnom |
100 for compressive force | ||||
Operating temperature |
oC |
−40 ... +120 | ||||
Relative reversibility error, 0.5 x Fnom |
% |
< 1 (typ. 0.5) | ||||
Relative linearity error |
% |
<± 1 (typ. 0.5) | ||||
Effect of temperature on output span / 10K |
% |
< 0.5 | ||||
Nominal (rated) displacement ("15 %) Insulation resistance Degree of protection per DIN EN 60529 |
μm Ω |
11 |
18 |
30 |
30 |
31 |
> 1013
IP65 | ||||||
Tightening torque for the connecting screws |
N⋅m |
0.5 |
1 |
2 |
4 |
21 |
Weight |
g |
8 |
22 |
137 |
240 |
720 |
Connection |
|
10−32 UNF |
Force Transducer PaceLine CFT - datasheet | PDF, 0.13 Kb |