KontrollVahend ja Tehnika (KVT) OÜ
Exclusive representative
Hottinger Baldwin Messtechnik GmbH, Germany
+372 6006190
+372 6006680


Products

Force Transducer PaceLine CLP

Special features

  • Symmetrical structure for improved linearity
  • Nominal (rated) forces: 7 kN, 26 kN and 62 kN
  • Rustless materials
  • Integrated charge cable

Type

CLP/...

Nominal (rated) force

Fnom

kN

7

26

62

Accuracy

Relative reversibility error

v

%

1

Relative linearity error

dlin

%

1

Electrical characteristics

Sensitivity (typical)1)

S

pC/N

-4.3

Insulation resistance

Ris

W

> 1013

Connection

Integrated cable, plug 10-32UNF

Temperature

Nominal (rated) temperature range

BT, nom

°C

-20 ¼ +120

Operating temperature range

BT, G

-20 ¼ +120

Storage temperature range

BT, S

-20 ¼ +120

Mechanical parameters

Max. operating force

FG

%

115

Limit force

FL

150

Breaking force

FB

200

Maximum bending moment with 2)

with Fz = 0% with Fz = 50% with Fz = 100%

M

Nm

0

5

0

0

35

0

0

134

0

Static lateral limit force 3)

FQ

% of Fnom

10

Nominal (rated) displacement

snom

mm

3

3.5

4.5

Fundamental resonance frequency

fG

kHz

105

120

140

Relative permissible oscillatory stress

Frb

% of Fnom

100

General information

Degree of protection per DIN 60529

IP65

Sensor material

Stainless steel, quartz

Cable sheath material

FPM (fluorinated rubber)

Potting material

Silicone

Cable length

m

0.5 or 1

Connector

10-32UNF

Ground

m

g

6.5

22

30

Specifications (data per VDI/VDE/DKD 2638 standards)b perm

1) Must be calibrated under mounting conditions

2) Fz is the force in the measurement direction

3) Related to a point of contact on the force application surface

Force Transducer PaceLine CLP - datasheet PDF, 0.2 Kb