Type |
C6A | ||||||
Accuracy class |
0.5 | ||||||
Nominal force |
Fnom |
MN |
200 kN |
500 kN |
1 MN |
2 MN |
5 MN |
Nominal sensitivity |
Cnom |
mV/V |
2 | ||||
Relative sensitivity deviation2) when used with hardened pressure plates when used with load button ZL and pendle bearing EPO3 when used with spherical cap ZK Relative zero signal deviation Relative range of inversion (0.5Fnom) Linearity deviation 2) when used with hardened pressure plates when used with load button ZL and pendle bearing EPO3 when used with spherical cap ZK Effect of temperature on sensitivity / 10 K by reference to nominal sensitivity Effect of temperature on zero signal / 10 K by reference to nominal sensitivity Creep over 30 min, in nominal temperature range 1) |
dc
ds,o u dlin
TKc TK0
dcrF+E |
% % % % %
% % %
%
%
% |
< ±2.5 < ±0.5 < ±2.5 < ±4 |
<±1 <±0.5 < ±0.5 | |||
<1 < ±0.8
< ±1 < ±0.5 < ±0.5 < ±0.5 < ±1 < ±0.5
< ±0.1
< ±0.05
< ±0.06 | |||||||
Input resistance at reference temperature Output resistance at reference temperature Isolation resistance at test voltage 100 V |
Re Ra Ris |
Ω Ω GΩ |
> 345 356 ±1.5 >5x 109 | ||||
Reference excitation voltage Operating range of the excitation voltage |
Uref BU,GT |
V V |
5 0.5 ... 12 | ||||
Reference temperature Nominal temperature range Operating temperature range Storage temperature range |
tref Bt,nom Bt,G Bt,S |
°C °C °C °C |
+23 ï10...+70 ï30...+85 ï50...+100 | ||||
Maximum operating force 1) |
FG |
% |
150 | ||||
Limit force 1) Breaking force 1) when used with hardened pressure plates when used with load button ZL and pendle bearing EPO3 when used with spherical cap ZK |
FL FB |
%
% % % |
150
>300 | ||||
>300 >200 |
>200 >200 | ||||||
Static lateral limit force 1) when used with hardened pressure plates when used with load button ZL and pendle bearing EPO3 when used with spherical cap ZK Permissible vibration amplitude 1) to DIN 50 100 |
FQ
Frb |
% % % % |
20 20 10 10 70 | ||||
Nominal displacement without mounting accessories ( ±15 %) Natural frequency without coupled ground and without mounting accessories Weight, without cable |
Snom fG |
mm kHz kg |
0.07
4.5
1.4 |
0.08
8
1.7 |
0.09
6
10.8 |
0.11
7.5
12.2 |
0.26
4.3
33 |
Degree of protection to DIN EN 60 529 Cable length, 6ïwire connection |
|
m |
IP 67 6 |
Force Transducer С6A - datasheet | PDF, 0.18 Kb |