KontrollVahend ja Tehnika (KVT) OÜ
Exclusive representative
Hottinger Baldwin Messtechnik GmbH, Germany
+372 6006190
+372 6006680


Products

Force Transducer U10M

Special features

  • Precise and robust tensile/compressive force transducer for static and dynamic measurement tasks
  • High lateral force and bending moment stability, the effect of the bending moment is electrically compensated
  • For forces up to 2.5 MN
  • The numerous possible configurations (TEDS, double bridge, various electrical connections, etc.), mean that it can be flexibly adapted to many measurement tasks
  • Made of rust-resistant materials, degree of protection IP68 on request
  • High fundamental frequency - ideal for measuring fast processes

Nominal (rated) force:

F

kN

1.25

2.5

5

12.5

25

50

125

250

500

MN

1.25

Accuracy

Accuracy class

0.02

0.03

0.04

0.05

Relative repro­ ducibility and re­ peatability errors without rotation

brg

%

0.02

Hysteresis error at

0.4 Fnom, rel. to the

full scale value

v0.4

%

0.02

0.03

0.04

0.05

Linearity deviation

dlin

%

0.02

0.025

0.035

0.05

Rel. zero point re­ turn

vw0

%

0.008

Relative creep

dcr, F+E

%

0.02

Effect of the bend­ ing moment at 10% Fnom * 10mm

dMb

%

0.01

Effect of lateral forces

(lateral force = 10% of Fnom)

dQ

%

0.01

Temperature coeffi­ cient of the rated output

TCS

% / 10 K

0.015

Temperature coeffi­ cient of zero signal

TC0

% / 10 K

0.015

Rated electrical output

Rated output (nomi­ nal)

Cnom

mV/V

1

2

Relative zero signal error

dS,0

%

1

Deviation of the rated output (with "adjusted rated out­ put" option)

dC

%

0.1

Rated output range (without "adjusted rated output" option)

C

mV/V

1 … 1.5

2 ... 2.5

Tension/compres­ sion rated output variation

dZD

%

0.2

Input resistance

Ri

Ω

>345

Output resistance (without "adjusted rated output" option)

Ro

Ω

280 ... 360

Output resistance (with "adjusted rated output" option)

Ro

Ω

365

280 ...

360

Tolerance of the output resistance in the "adjusted rated output" option

dRa

%

±0.5 Ω

-

Insulation resis­ tance

Ris

>2

Operating range of the excitation voltage

BU,G

V

0.5 … 12

Reference excita­ tion voltage

Uref

V

5

Connection

6-wire circuit

Force Transducer U10M - datasheet PDF, 0.21 Kb