Type |
U1A | ||||
Accuracy class |
0.1 | ||||
Nominal force |
N |
10 |
20 |
50 | |
Nominal sensitivity rel. sensitivity deviation rel. tensile-compressive sensitivity difference rel. deviation from zero Rel. range of inversion (0.2Fnom to Fnom) Linearity deviation |
Cnom dc dzd ds,o u dlin |
mV/V % % % % % |
2 0.2 0.4 1 0.15 0.1 | ||
Temperature effect on the sensitivity/10 K by reference to nominal sensitivity Temperature effect on the zero signal/10 K by reference to nominal sensitivity Effect of eccentricity / mm Rel. creep over 30 min |
TKC
TK0 dE dcrF+E |
%
% % % |
0.1
0.05 0.1 ±0.06 | ||
Input resistance Output resistance Isolation resistance |
Re Ra Ris |
Ω Ω Ω |
>345 300 − 400 >5 x 109 | ||
Reference excitation voltage Operating range of the excitation voltage |
Uref BU,G |
V V |
5 0.5 ... 6 | ||
Nominal temperature range Operating temperature range Storage temperature range Reference temperature |
Bt,nom Bt,G Bt,S tref |
oC oC oC oC |
−10 ... +50 −20 ... +50 −30 ... +60 +23 | ||
Max. operational force Limit force Breaking force Static lateral limit force |
(FG) (FL) (FB) (FQ) |
% % % % |
120 | ||
1000 2000 200 |
500 1000 100 |
200 400 40 | |||
Nominal displacement Fundamental resonance frequency Weight Rel. permissible vibrational stress Degree of protection to DIN EN 60529 |
Snom fG
Frb |
mm Hz kg % |
0.28 300 |
0.21 450 |
0.2 700 |
0.3 70 IP 40 |
Force Transducer U1A - datasheet | PDF, 0.17 Kb |