Type |
Z4A | ||||||
Data according to VDI standards 2638 |
| ||||||
Nominal (rated) force |
Fnom |
kN |
20 |
50 |
100 |
200 |
500 |
Accuracy class1) |
|
00 |
0.5 | ||||
Nominal sensitivity rel. sensitivity deviation rel. tensile/compressive force sensitivity diff. Zero signal tolerance rel. zero point compensation (zero signal return)1) |
Cnom dc dzd ds,o fo |
mV/V % % % % |
2 0.1 0.2 0.5 0.08 | ||||
Rel. range (0.2Fnom to Fnom) at:1) unchanged mounting position, typically different mounting positions, typically Rel. range of inversion (0.2Fnom − Fnom)1) |
bi b u |
% % % |
0.02 0.03 | ||||
0.06 |
0.15 | ||||||
Linearity deviation |
dlin |
% |
0.02 |
0.03 | |||
Effect of temperature on sensitivity/10 K by reference to nominal sensitivity
Effect of temperature on zero signal/10 K by reference to nominal sensitivity |
TKc TK0 |
%
% |
0.01
0.015 | ||||
Effect of transverse forces (Transverse forces 10 % Fnom)2) |
dQ |
% |
0.03 | ||||
Effect of eccentricity per mm Rel. creep over 30 min. |
dE dcrF+E |
% % |
0.01 |
0.005 | |||
0.02 | |||||||
Input resistance Output resistance Isolation resistance |
Re Ra Ris |
Ω Ω Ω |
>345 356±0.3 >5⋅109 | ||||
Reference excitation voltage Operating range of the excitation voltage |
Uref BU,G |
V V |
5 0.5 ... 12 | ||||
Nominal temperature range Operating temperature range Storage temperature range Reference temperature |
Bt,nom Bt,G Bt,S tref |
°C °C °C °C |
+10...+40 -30...+85 -50...+85 +22 | ||||
Max. operational force Limit force Breaking force Static lateral limit force |
(FG) (FL) (FB) (FQ) |
% % % % |
150 150 250 30 | ||||
Limit torque |
MG |
N⋅m |
120 |
350 |
950 |
2000 |
4000 |
Nominal displacement |
Snom |
mm |
0.2 |
0.25 |
0.28 |
0.45 | |
Fundamental resonance frequency |
fG |
kHz |
4.1 |
4.5 |
3.4 |
3.6 |
2.5 |
Weight |
|
kg |
1.8 |
2.4 |
5.5 |
11.2 |
42 |
Rel. permissible vibrational stress |
Frb |
% |
70 |
50 | |||
Cable length, six−wire connection |
m |
6 | |||||
Degree of protection to DIN EN60529 |
IP 67 |
Force Transducer Z4A - datasheet | PDF, 0.2 Kb |