Type |
K3607 | |
Accuracy class |
mV/V |
0.025** |
Calibration steps |
| |
5 range steps |
0.5; 1; 2; 5; 10 | |
11 percentage steps within the range in each case |
% |
0; 10; 20...100 |
Polarity switch* |
|
positive or negative output signal as disired |
Step error of the range steps, |
|
|
related to the full scale value of the range in each case |
% |
< ± 0,025 |
Step error of the percentage steps, |
|
|
related to the full scale value of the range in each case |
% |
< ± 0,01 |
Absolute calibration of th full scale value |
|
|
of the range et the 2 mV/V at 23 °C |
|
|
with 6-wire cable, related to 2 mV/V |
% |
< ± 0,02 |
Temperature effect of the absolute calibration per 10 K, |
|
|
in the service temperature range |
% |
< 0,03 |
in the nominal temperature range |
% |
< 0,01 |
Nominal temperature range |
°C |
+10...+50 |
Permissible ambient temperature range |
°C |
+10...+60 |
Service temperature range |
°C |
+10...+60 |
Storage temperature range |
°C |
-25...+70 |
Frequency range of supply voltage |
Hz |
0(DC)...5000 Hz** |
Nominal value of supply range |
V |
4 |
Maximum permissible supply voltage |
V |
18 |
Weight |
kg |
1 |
Dimensions (b x h x t) |
mm |
200 x 110 x 120 |
Calibration Unit K3607 - datasheet | PDF, 0.09 Kb |