Type |
K800 | ||||
Calibration resistance |
Ω |
120 |
350 |
700 |
1000 |
Accuracy class |
|
0.03 | |||
Nominal (rated) value of the excitation voltage Maximum permissible excitation voltage |
VDC VDC |
2.5 5 | |||
Calibration steps
Equivalence (at gage factor = 2) |
% mV/V µm/m |
−100, −50, 0, 50, 100 −8, −4, 0, 4, 8 −16000, −8000, 0 , 8000, 16000 | |||
Range deviation Influence of temperature on absolute calibration per 10K, in nominal temperature range |
%
% |
< 0.03
< 0.03 | |||
Nominal temperature range Operating temperature range Storage temperature range |
_C _C _C |
+10...+40 0...+60 –25...+70 | |||
Supply voltage AP814 connection AP815 connection |
V |
±12 or via USB 3-wire technique, 1x 25-pin SubD 4-wire technique, 2x 25-pin SubD | |||
Dimensions (H x W x D) K800 calibration instrument Power supply unit Weight, approx. K800 calibration instrument Power supply unit |
mm mm
kg kg |
75 x 330 x 270 60 x 120 x 65
3 0.5 |
Calibration unit for 8-channel quarter-bridge measuring amplifiers K800 - datasheet | PDF, 0.22 Kb |