KontrollVahend ja Tehnika (KVT) OÜ
Exclusive representative
Hottinger Baldwin Messtechnik GmbH, Germany
+372 6006190
+372 6006680


Products

Digital precision measuring instrument DMP41

Special features

  • Accuracy class 0.0005
  • Simultaneous measurement with optional 2 or 6 channels
  • Resolution of measured signals up to the physical limit > 1,000,000 d
  • Individual linearization
  • Powerful digital filters
  • Background calibration
  • Operation via touchscreen, keyboard or computer
  • Improved temperature and EMC behavior
  • LAN/USB as required

Type

DMP41−T2

DMP41−T6

Accuracy class

0.00051)

Number of amplifiers

Transducers that can be connected

2

2 SG full bridges

6

6 SG full bridges

Excitation voltage (−) Ub Carrier frequency

Transducer resistance per amplifier

at Ub = 2.5V at 5V

at Ub = 10V

Length of transducer cable Measuring ranges

Digital filter (6th order) Display resolution Common-mode rejection Input resistance

V

Hz

W

W

W

m mV/V Hz digit dB MW

2.5; 5; 10

225 ±100ppm

75 … 4,000

150 … 4,000

300 … 4,000

< 200

±2.5; ±5, ±10 40...0.01 (15 steps)

> 1,000,000

> 120

> 100

Sampling rate, per amplifier

1/s

1 … 450

Taring range/Zeroing range

Linearization of transducer characteristic curve

Total display range 2...11 points

Non-linearity

rel. to full scale value

Temperature effect per 10 K in nominal (rated) temperature range

on zero point (rel. to full scale value) on sensitivity (rel. to actual value)

Short-term drift over 5 min, from 2 h after switch-on

Long-term drift over 24 h, from 2 h after switch-on

%

%

%

ppm ppm

< 0.0005

< 0.0002

< 0.0005

max. ±2, typ. ±1 max. ±5, typ. ±2

Nominal (rated) temperature range Operating temperature range Storage temperature range

°C

°C

°C

10...+40

10...+50

-10...+60

Operating voltage (mains voltage)

Power consumption Weight, (net weight) Dimensions (W x H x D)

V W

kg

mm

85 … 264 (50...60 Hz)

approx. 35 approx. 45

approx. 9 approx. 9,5

458 x 171 x 367

Connections for

SD transducer in six-wire configuration

Temperature sensor (1-wire), max. 4 sensors Digital inputs and outputs

Computer interface Ethernet

Computer interface USB USB host interface

Computer interface serial (optional)

2 x D-SUB-15 6 x D-SUB-15

2 x Amphenol 6 x Amphenol

RJ45

D-SUB-15 RJ45

USB device 2 x USB Host

Adapter D-SUB-9

Range of use

Indoor

Altitude, max.

m

2000

Protection class

I

Overvoltage category

II

Permitted degree of contamination

2

Rel. air humidity, max.

80 % at 31 oC, decreasing linearly to 50% at 40oC

Degree of protection

IP 20 per DIN EN 60529

Digital precision measuring instrument DMP41 - datasheet PDF, 0.15 Kb