digiCLIP |
| ||
Accuracy class (at UB = 2.5 V and UB = 1 V); after autocalibration |
|
0.05, typically 1.1 in an industrial environment as per EN 61326 1.2 in the 10 mV/V measuring range | |
Power supply | |||
Supply voltage, Overvoltage and reverse polarity protection |
VDC |
24 | |
Isolation voltage Potential separation between the supply bus and the transducer connection, functional sep- aration, must not be used for safety consider- ations |
VDC |
500 | |
Permissible supply voltage range |
V |
18 ... 30 | |
Influence of supply voltage when there are changes in the specified range |
%/V |
< 0.001 | |
Power consumption, max.; incl. transducer |
W |
1.5 | |
Amplifier | |||
Carrier frequency, square |
Hz |
600 (591.9 Hz ±100 ppm) | |
Synchronization |
|
when several interconnected modules are used, the carrier frequency is synchronized automatically | |
Bridge excitation voltage UB, Peak−to−peak (±10%) |
V |
2.5 |
1.0 |
Measuring range |
mV/V |
±4 |
±10 |
Connectable transducers SG full bridge |
ohms |
80 ... 5000 | |
Connection technique |
|
4 and 6−wire circuitry with single−wire open−circuit monitoring | |
Permissible cable length between transducer and amplifier, max. |
m |
100 | |
Input resistance |
MOhm |
>5 | |
Measurement frequency range, adjustable (−3dB) (see filter table) |
Hz |
0.05 ... 225 | |
Filter characteristics |
|
Bessel, 4th order | |
Noise voltage relative to input, for UB = 2.5 V, typical |
µV/V |
1.0 (at 100 Hz filter frequency) 0.05 (at 1 Hz filter frequency) | |
Influence of ambient temperature for change of 10 K on the zero point (TK0) on sensitivity (TKC) |
µV/V % |
0.1 0.05 f.s. | |
Linearity deviation |
% f.s. |
0.005 | |
Long−term drift, without AutoCal |
% |
<0.001 (within 48 h) | |
Communication interface | |||
Number of devices on the bus, max. Address settings Protocol
Hardware bus link Bit rate Line length, max. Bit rate selection PDO transfer Cycle time for time−driven triggering, Possibly restricted by chosen data types and filter frequency 1) DeviceNet connection |
kBit/s m
ms |
64 0 to 63 via rotary switch on front DeviceNet standard: “The DeviceNet Specification”, available at www.odva.org Two-wire, as per ISO 11898 500 250 125 100 250 500 Automatic recognition after change of address Triggered by sampling rate, timing control or SYNC message
0.85 ... 25000
Plug terminal on the side: potential separation from power supply and measurement ground. Option: DF001: 9−pin sub-D (DIN 19245) | |
Signal conditioning | |||
A/D converter |
|
Delta-Sigma, 24−bit | |
Scaling accuracy |
bits |
32 | |
Sampling rate |
1/s |
1184 |
Digital amplifier digiCLIP DF30DN - datasheet | PDF, 0.34 Kb |