Type |
SCOUT 55 | |
Accuracy class |
0.1 | |
Mains connection / Supply voltage |
V |
115/230, +6 %; -14 %; |
|
Hz |
48 ¼ 60 |
Power consumption, max. |
VA |
8 |
Safety fuse (slow blowing) |
mA |
T 125 mA L (115 V) / T 63 mA L (230 V) |
Amplifier Carrier frequency Excitation voltage UB (±5 %) Transducers that can be connected S.G. half and full bridge Inductive half and full bridge, LVDT's
Permissible cable length between transducer and am- plifier Measurement frequency range, adjustable (-1 dB) Input level Measuring range UB=2.5 V UB=1 V Bridge balance range UB=2.5 V UB=1 V Noise voltage1) 0...200 Hz 0...1.25 Hz Effect of 10 K change in ambient temperature1) Autocalibration on / off Sensitivity Zero point |
Hz Vrms
W mH
m Hz mV/V mV/V mV/V mV/V mV/VPP mV/VPP
% mV/V |
4800±0.32 1 or 2.5 UB = 1 Vrms 40 ¼ 5000 6 ¼ 19 UB = 2.5 Vrms 80 ¼ 5000 2.5 ¼ 20
max. 500
0.05 ¼ 1000 low medium high 0.2 ¼ 4 2 ¼ 40 20 ¼ 400 0.5 ¼ 10 5 ¼ 100 50 ¼ 1000
±4 ±40 ±400 ±10 ±100 ±1000
0.5 1 10 0.025 0.1 1
0.04 / 0.1 0.04 / 0.1 0.04/0.1 0.2/2 2/20 20/200 |
Measurement frequency range |
|
Nom. val. fc ‐1 dB ‐3 dB Phase del. Rise time Overshoot |
Butterworth low‐pass |
(Hz) (Hz) (Hz) (ms) (ms) (%) | |
|
1000 1010 1165 0.66 0.35 12 | |
|
500 485 580 1.1 0.7 12 | |
|
200 245 290 1.7 1.3 11 | |
|
80 78 98 4.3 3.8 10 | |
|
40 38 50 7.1 7.3 8 | |
|
20 19 26 12 14 7 | |
|
10 9.1 12.5 22 28 6 | |
|
5 4.6 6.3 41 56 5 | |
Bessel low pass |
Nom. val. fc ‐1 dB ‐3 dB Phase del.Rise time Overshoot | |
(Hz) (Hz) (Hz) (ms) (ms) (%) | ||
900 900 1550 0.49 0.28 4.1 | ||
400 400 750 0.8 0.6 2 | ||
200 215 395 1.3 1.0 2 | ||
100 111 190 2.5 2.1 2.5 | ||
40 39 68 5 5.5 1.1 | ||
20 21 37 8.1 10 1 | ||
10 11 19 14 19 0.7 | ||
5 5.3 9.7 25 38 0.3 | ||
2.5 2.7 4.9 48 75 0 | ||
1.25 1.4 2.4 90 150 0 | ||
0.5 0.7 1.2 180 300 0 | ||
0.2 0.17 0.3 700 1200 0 | ||
0.1 0.09 0.16 1400 2300 0 | ||
0.05 0.044 0.075 2900 4700 0 | ||
Max. permissible common‐mode voltage |
V |
±5 V |
Common‐mode rejection |
dB |
typically 110 |
Max. differential voltage DC |
V |
±10 |
Linearity deviation |
% |
typically 0.05 |
Long‐term drift over 48 hours, Meas. range 2 mV/V |
|
Autocalibration on / off |
30 minutes after switching on (warm‐up time) |
mV/V |
<0.2 / <0.4 |
Measuring Amplifier SCOUT55 - datasheet | PDF, 0.33 Kb |