The MGCplus system is tested according to the harmonized European standards EN61326 and EN61010 Therefore it achieves the protection aims of EEC directive 89/336EEC in the field of electromagnetic compatibility and EEC directive 73/23/EEC in the field of electrical safety of low voltage equipment. Mechanical stress is tested according to European Standard EN60068-2-6 for vibrations and EN60068-2-27 for shock. The equipment is exposed to an acceleration of 25 m/s2 within the frequency range of 5...65 Hz in all 3 axes. Duration of this vibration test: 30 min per axis. The shock test is performed with a rated value of 200 m/s2 acceleration during 11 ms, half sine pulse shape, with 3 shocks in each of the 6 possible directions.

The maximum capacity of drawn current on each MGCplus slot is 150 mA in a 16 slots housing. The double capacity on every used slot is possible, if the next slot of each used one is left free.

General technical data single-channel amplifier modules

Width

Maximum sampling rate without linearisation

mm values/s

20.3 (4 divs)

19200

Limit value switches

%

4

Number

Reference level

Gross, net, peak values

Reference voltage (freely adjustable)

−100 ...+100 of measuring range

Hysteresis factory setting

%

1 of measuring range

Adjustment accuracy

%

0.0033 of measuring range

Response time

ms

1.0

with Butterworth filter > 5 Hz

and Bessel filter > 1.25 Hz

Peak value store

μs

2

Number

Function

Maximum, minimum

Peak combine

Peak-peak, average

Update time

30

with Butterworth filter > 250 Hz

and Bessel filter > 100 Hz

Clearing the peak value stores (switching to current value)

within 1 ms, via control inputs

Holding of the current value/peak value

within 1 ms, via control inputs

Time constant for envelope curve function

s

0.01 ... 10000

Control contacts (HCMOS)

V

−0.5 to +5.5

Inputs (8 lines, connectable as desired)

Permissible input voltage

High level

V

min. +4.0

Low level

V

max. +0.7

Schmitt-trigger, hysteresis

V

> 1.1

Pull-up resistors

100

Outputs (limit value switch, error)

V

> 4.0

High level with max. 1 mA

Low level with max. 0.7 mA

V

< 0.7

Internal resistance

1

Nominal (rated) temperature range

°C [°F]

−20 ...+60 [−4 ... +140]

Service temperature range

°C [°F]

−20 ...+60 [−4 ... +140]

Storage temperature range

°C [°F]

−25 ...+70 [−13 ... +158]

Operating voltages

V

±14.6...17.0; (< 120 mA) ±7.7...8.3; (< 120 mA)2)

+4.9...5.1; (< 150 mA)

Card format (Euro)

mm

100 x 160

Weight

g

300

Connection plug

IEC 603-2 indirectly

Analog outputs Vo1 and Vo2

Nominal (rated) voltage

V

±10 V (asymmetrical)

Permissible load resistance

> 5

Internal resistance

Ω

< 5

The two output voltages can be used to display the five

Amplifier output with zero balance

signal voltages.

Output staggered by the tare value

Peak value store 1 output (maximum or minimum)

Peak value store 2 output (maximum or minimum)

Peak value store 3 output (peak combine of 1 and 2)

Max. deviation of the analog outputs from the

mV

< 3 (< 10 with ML10B)

digital value

Monitor output for Vo1 via BNC plug on front panel

Nominal (rated) voltage

V

±10 (asymmetrical)

Permissible load resistance

> 1000

Internal resistance

1

Effect of a 10K-change of the ambient temperature

(effect additional to digital value)

on the outputs Vo1 and Vo2:

sensitivity

%

< 0.08, typ. 0.04

zero point

mV

< 3 typ. 2