KontrollVahend ja Tehnika (KVT) OÜ
Exclusive representative
Hottinger Baldwin Messtechnik GmbH, Germany
+372 6006190
+372 6006680


Products

QuantumX MX1615B

Special features

  • 16 individually configurable inputs
  • Connection of strain gauges in quarter- , half- and full-bridge
  • Bridge excitation : DC or carrier frequency
  • Internal shunt resistor
  • Connection of standard voltage, PT100, resistor, Potentiometer
  • Individual data rates up to 19,200 Hz per channel, active low pass filter
  • 3000 Hz bandwidth
  • 24−bit A/D converter per channel for synchronous, parallel measurements

General specifications

Inputs

Number

16, electrically isolated from the supply

Transducer technologies, can be adjusted individually

Strain gauges in full- , half- or quarter-bridge configuration. Selectable bridge excitation voltage : DC voltage or carrier frequency (AC/CF, 1200 Hz)

SG-quarter bridges SG-half bridges SG-full bridges

Three wire and four wire five wire

six wire

Resistor, Resistance thermometer (PT100), connection in four-wire configuration

Potentiometric transducers

Voltage (±10 V differential, 0 … 30 V unipolar) without transducer supply

A/D converter per channel

24 Bit Delta Sigma converter

Data rates

Hz

0.1 … 19,200 adjustable for each channel

Bandwidth

kHz

3.0

0.4 using carrier frequency

Active low-pass filter (Bessel/Butterworth, can be switched off)

Hz

0.01 … 3,000

Transducer identification (TEDS, IEEE 1451.4)

max. distance of the TEDS module

m

100

Transducer connection

Phönix Contact FMC−1,5/8−ST−3,5−RF plug terminal Plug included

Supply voltage range (DC)

V

10 ... 30 (24 V nominal (rated) voltage)

Supply voltage interruption

max. 5 ms at 24 V

Power consumption

W

< 12

Ethernet (data link)

Protocol/addressing Connection

Max. cable length to module

m

10Base-T / 100Base-TX

TCP/IP (direct IP address or DHCP)

8P8C plug (RJ-45) with twisted pair cable (CAT-5) 100

Synchronization options

EtherCATâ IEEE1588v2 (PTP), NTP

IRIG−B (B000 to B007; B120 to B127)

IEEE1394b FireWire (only QuantumX, automatically, recommended)

via CX27 via Ethernet

via MX440A - or MX840A input channel

IEEE1394b FireWire (module synchronization, data link, optional supply voltage)

Baud rate

Max. current from module to module Max. cable length between the nodes

Max. number of modules connected in series (daisy chain)

Max. number of modules in a IEEE1394b FireWire system (including hubs, backplane)

Max. number of hops

MBaud A

m

IEEE 1394b (HBM modules only)

400 (approx. 50 MByte/s)

1.5

5

12 (=11 hops)

24

14

Nominal (rated) temperature range

°C [°F]

−20 ... +65 [−4 ... +149]

Storage temperature range

°C [°F]

−40 ... +75 [−40 ... +167]

Rel. humidity

%

5 ... 95 (non condensing)

Protection class

III

Degree of protection

IP20 per EN 60529

Mechanical tests

Vibration (30 min)

Shock (6 ms)

m/s2 m/s2

50

350

EMC requirements

per EN 61326−1

Max. input voltage at transducer socket to ground, transient free

Pin 6 and 7 to Pin 1, 2, 3, 4 or 5

V

±18

Dimensions, horizontal (W x H x D)

mm

52.5 x 200 x 122 (with case protection)

44 x 174 x 119 (without case protection)

Weight, approx.

g

980

QuantumX MX1615B - datasheet PDF, 0.26 Kb