General specifications | ||
Inputs |
Number |
2, electrically isolated from each other and from the supply voltage |
Transducer technologies |
|
Strain gauge full bridge |
A/D converter |
|
24 Bit Delta Sigma converter |
Signal bandbwidth |
Hz |
50 |
Aktive low-pass filter |
Hz |
Bessel, Butterworth, 0.01 ... 50 |
Sample rate (Data output rate adjustable by software, default: HBM Classic) |
S/s |
Dezimal: 0,1 …40.000 HBM Classic: 0,1 … 9.600 |
Transducer identification (TEDS, IEEE 1451.4) max. distance of the TEDS module |
m |
100 |
Transducer connection |
|
D-SUB-15HD |
Supply voltage range (DC) |
V |
10 ... 30 (24 V nominal (rated) voltage) |
Supply voltage interruption |
|
max. 5 ms at 24 V |
Power consumption |
W |
< 7 |
without adjustable transducer excitation | ||
with adjustable transducer excitation |
W |
< 10 |
Transducer Excitation (active transducers) Adjustable supply voltage (DC) Maximum output power |
V W |
5 ... 24; adjustable for each channel 0.7 each channel / a total of 2 |
Ethernet (data link) Protocol/addressing Connection
Max. cable length to module |
− −
m |
10Base-T / 100Base-TX TCP/IP (direct IP address or DHCP) 8P8C plug (RJ-45) with twisted pair cable, Streaming (CAT-5) 100 |
Synchronization options
EtherCATâ IRIG−B (B000 bis B007; B120 bis B127) IEEE1588 (PTPv2), NTP |
|
IEEE1394b FireWire (only QuantumX, automatically, recommended) via CX27 EtherCAT Gateway via MX440A/B- or MX840A/B-input channel Ethernet based Network Time Protocol |
IEEE1394b FireWire (module synchronization, data link, |
MBaud |
IEEE 1394b (HBM modules only) |
optional supply voltage) |
| |
Baud rate |
400 (approx. 50 MByte/s) | |
Max. current from module to module |
A |
1.5 |
Max. cable length between the nodes |
m |
5 |
Max. number of modules connected in series (daisy chain) |
− |
12 (=11 Hops) |
Max. number of modules in a IEEE1394b FireWire system |
|
|
(including hubs, backplane) |
− |
24 |
Max. number of hops |
− |
14 |
Nominal (rated) temperature range |
°C [°F] |
−20 ... +65 [−4 ... +149] |
Storage temperature range |
°C [°F] |
−40 ... +75 [−40 ... +167] |
Rel. humidity |
% |
5 ... 95 (non condensing) |
Protection class |
|
III |
Degree of protection |
|
IP20 per EN 60529 |
Mechanical tests |
m/s2 |
50 |
Vibration (30 min) | ||
Shock (6 ms) |
m/s2 |
350 |
QuantumX MX238B - datasheet | PDF, 0.3 Kb |