KontrollVahend ja Tehnika (KVT) OÜ
Exclusive representative
Hottinger Baldwin Messtechnik GmbH, Germany
+372 6006190
+372 6006680


Products

Beam Load Cell HLC

Special features

  • Hermetically sealed (IP68)
  • Max. capacities: 220 kg ... 10 t
  • Stainless steel
  • Low overall height
  • Meets EMC/ESD requirements according to EN 45 501
  • Complies with OIML R60 regulations up to 6000d
  • Explosion‐proof versions according to ATEX 95 optional

Type HLC A1 ...

Maximum capacity (Emax)

(Load introduction = thread through)

HLC A1 D1 / ... + HLC A1 C3 / ...

220 kg; 550 kg; 1.1 t; 1.76 t; 2.2 t; 4.4 t

Type HLC B1 ...

Maximum capacity (Emax)

(Load introduction = counterbore + thread) 3)

HLC B1 D1 / ...

220 kg; 550 kg; 1.1 t; 1.76 t; 2.2 t; 4.4 t; 10 t

HLC B1 C3 / ...

220 kg; 550 kg; 1.1 t; 1.76 t; 2.2 t; 4.4 t

HLC B1 C4 / ... + HLC B1 C6 / ...

220 kg; 550 kg; 1.1 t

Type HLC F1 ...

Maximum capacity (Emax)

(Load introduction = blind hole + integrated overload stop)

HLC F1 D1 / ... + HLC F1 C3 / ...

220 kg; 550 kg; 1.1 t; 1.76 t

Accuracy class according to OIML R60 Maximum number of load cell intervals (nLC)

D1

1000

C3

3000

C4 5)

4000

C6 5)

6000

Minimum LC verification interval (vmin)

%

of Emax

0,0285

0.0100 (220kg; 1.76 t; 2.2 t; 4.4 t)

0.0090 (550 kg + 1.1 t)

Sensitivity (CN)

mV/V

1.94 (10 t = 2.00 mV/V)

Sensitivity tolerance

%

±0.5

±0.1

Temperature effect on zero balance (TK0)

% of Cn

/ 10 K

±0.0400

±0.0140 (220 kg; 1.76 t; 2.2 t; 4.4 t)

±0.0126 (550 kg + 1.1 t)

Temperature effect on sensitivity (TKC) 4)

±0.0420

±0.0140

±0.0105

±0.0070

Hysteresis error (dhy) 4)

% of Cn

±0.0500

±0.0166

±0.0125

±0.0083

Non‐linearity (dlin) 4)

±0.0500

±0.0170

±0.0166

Creep (dcr) over 30 min.

±0.0500

±0.0166

±0.0166

±0.0122

Minimum dead load output return (MDLOR)

±0.0500

±0.0166

±0.0125

±0.0083

Input resistance (RLC)

Ω

350 ... 480

Output resistance (R0)

350 ±2

350 ±0.12

Reference excitation voltage (Uref)

V

5

Nominal range of excitation voltage (BU)

0.5 ... 15

( Ex-Versions max. 12 V !!! )

5 ... 10

Insulation resistance (Ris)

> 5

Nominal temperature range (BT)

°C [°F]

-10 ... +40 [+14 ... +104]

Service temperature range (Btu)

-30 ... +70 [-22 ... +158]

Storage temperature range (Btl)

-50 ... +85 [-58 ... +185]

Safe load limit (EL)

%

of Emax

150

Lateral load limit (Elq)

100

Breaking load (Ed)

300

Permissible dynamic load (Fsrel)

(vibration amplitude according to DIN 50100)

70

Deflection at Emax (snom), approx.

mm

0.5 (1.76 t = 1.4 mm)

Weight (G), approx.

kg

0.9 (220 kg ... 1.76 t); 1.6 (2.2 t); 2.2 (4.4 t); 6.2 (10 t)

Protection class to EN 60 529 (IEC 529)

IP68

Material: Measuring element Cable fitting Cable‐sheath

Stainless steel 6)

Stainless steel 6) / Gasket: Viton®

PVC

1) Maximum capacity 10 t: Load introduction = counterbore + hole

2) The data for Non‐linearity (dlin), Hysteresis error (dhy) and Temperature effect on sensitivity (TKC) are typical values. The sum of these data meets the requirements according to OIML R60.

3) Accuracy classes C4 and C6: HLC B1 ... / 220 kg; 550 kg; 1,1 t only

4) According to EN 10088-1

Beam Load Cell HLC - datasheet PDF, 0.22 Kb