Type |
FIT5A | ||||||
Accuracy class 1) |
|
C3 3000 | |||||
Number of load cell verification intervals |
nLC |
| |||||
Fraction |
Plc |
|
0.8 | ||||
Maximum capacity |
Emax |
kg |
5 |
10 |
20 |
30 |
50 |
Minimum load cell verification interval (standard) |
vmin |
g |
0.5 |
1 |
2 |
5 |
5 |
Minimum load cell verification interval (option) |
vmin |
0.2 |
0.5 |
1 |
2 |
2 | |
Temperature coefficient of the zero signal per 10 K (standard) 1)2) |
TC0 |
% of Cn |
±0.0160 |
0.0260 |
0.0160 | ||
Temperature coefficient of the zero signal per 10 K (option) 1)2) |
TC0 |
% of Cn |
±0.0064 |
±0.0080 |
±0.0106 |
±0.064 | |
Maximum platform size |
|
mm |
400 x 400 | ||||
Nominal (rated) sensitivity |
Cn |
digit |
1 000 000 | ||||
Zero signal |
|
0 ±100 000 | |||||
Temperature coefficient of the sensitivity per 10 K in the temperature range 4) +20 … +40°C −10 … +20°C |
TCS |
% of Cn |
±0.0218 ±0.0150 | ||||
Non-linearity 4) |
dlin |
±0.0166 | |||||
Relative reversibility error 4) |
dhy |
±0.0166 | |||||
Minimum dead load output return |
MDLOR |
±0.0166 | |||||
Off-center load error 3) |
|
±0.0166 | |||||
Nominal (rated) ambient temperature range |
BT |
°C |
−10 … +40 | ||||
Operating temperature range |
Btu |
−10 … +50 | |||||
Storage temperature range |
Btl |
−25 … +70 | |||||
Limit load at 20 mm eccentricity |
EL |
% of Emax |
1000 | ||||
Service load at max. 120 mm eccentricity |
|
150 | |||||
Relative vibrational stress at max. 50 mm eccentricity |
Fsrel |
70 | |||||
Nominal (rated) displacement 5) |
snom |
mm |
<0.2 | ||||
Weight, approx. |
G |
kg |
3 | ||||
Degree of protection per DIN EN 60 529 (IEC 529) |
|
|
IP68 (test condition 1 m water column / 100 h); IP69K (water at high pressure, steam cleaner) 6) | ||||
Voltage supply Operating voltage (DC) Power consumption Start-up current |
UB |
V W A |
+10 … +30 :::2 <0.2 | ||||
Measuring body material |
|
|
Stainless steel 1.4545 7) |